ZH300 NDIR Infrared Gas Analyzer

ZH300 NDIR Infrared Gas Analyzer

Analyzer Characteristics

According to the Lambert-Beer law, the characteristic absorption intensity is proportional to the gas concentration. It is an NDIR (non-dispersive) infrared gas analyzer, which can be used to continuously analyze the concentration of one or several gas components in the mixed gas.

 

The SJ300 infrared gas analyzer adopts the most mature and reliable analysis method in the field of gas analysis, and selects the most advanced MEMS infrared light source and dual-channel infrared detector in the world.The SJ300 infrared gas analyzer has complete functions and superior performance indicators, especially good stability, strong anti-interference ability, little influence by ambient temperature and high reliability. It is suitable for online use in harsh environment process industries, environmental protection, and scientific research fields.

Technical characteristics
Project
Indicators
CO Detection range
Min: 0 - 50 ppm, Max: 0 - 100%vol
CO2 Detection range
Min: 0 - 50 ppm, Max: 0 - 100%vol
CH4 Detection range
Min: 0 - 2000 ppm, Max: 0 - 100%vol
CH4: 0~100%; CnHm: 0~10%; CO 0~50%; CO2: 0~20%;H2: 0~10%
Calorific value Detection range
O2: 0-25%. ECD/paramagnetic oxygen
SF6 Detection range
0 - 100 ppm
Linearity error
≤ ±2% F.S.
Repeatability
≤ 1%
Zero drift
≤ ±2% F.S./ week
Span drift
≤ ±2% F.S./week
Response time (T90)
T90≤30s
Size
19″x 3U x 320 mm
Signal transmission
RS485,4-20mA
Power
220 V AC/50 Hz 100 W